Complete workflow
The full session: stimulation parameters, electrode configuration, scale ratings and notes, recorded configuration by configuration.
The screen is a four-step wizard. Steps 2 and 4 share a two-row layout that follows the order the work is actually done in: what was delivered on top (parameters and electrodes), what was observed below (scales and notes).
A visit from the home screen: a new file, the stimulation parameters, the lead, the clinical scales, then three configurations rated and inserted, with the charts and table below.
Step 1: File
The File step, with an existing session opened.
Enter the patient ID, the task and the run number, then choose where
to save. The task is programming unless you name another; whatever you type
becomes the task- part of the BIDS filename and
the TaskName in its sidecar, so a study that labels its tasks differently can
keep its own scheme. Like every BIDS label it keeps letters and digits only.
New asks whether this visit is a loose TSV or goes straight into a BIDS dataset.
A loose TSV goes wherever you choose. The app composes the
BIDS filename and creates the file straight away,
along with its .json sidecar, so every later insert has somewhere to go.
Into a dataset asks for the dataset folder, which must be a BIDS dataset or
an empty folder to start one in (see
what counts as a BIDS dataset). If the folder
you picked is neither, Choose another folder opens the picker again. It then
asks for the session label, proposed as today’s date but editable:
ses-YYYYMMDD is this app’s convention, and plenty of studies use
ses-preop or ses-3mo. As you type, the dialog shows the label as it will
be filed (3-mo becomes 3mo).
A dataset can hold any number of recordings per participant and session:
several ses-preop visits, programming and notes side by side. The one thing
two files cannot share is the whole name, participant, session, task and run
together, because the second would replace the first on disk. So when the name
this visit would get is already in the dataset, the dialog says so and the visit
is filed under the next free run instead (run-02, run-03), and the
Run field changes to match. Nothing is refused and nothing is overwritten.
The label you choose is also the one used when you export this visit. The visit is
filed at its BIDS path when the first block is inserted, not when the file
is created: a dataset holding a header-only TSV that no scans.tsv lists is
one that does not validate, and before the first block there is nothing to file.
From then on every insert rewrites it in place, and the dataset index files stay
consistent. It goes through the same rules as
adding to a dataset, so your
dataset_description.json and participants.tsv columns (the dataset’s
own description and participant list) are safe. This option is
available on Windows, macOS and Linux. On iPadOS and Android, which do not let an
app write into a folder you pick, the choice still appears with Into a
dataset greyed out and the reason beneath it: record a loose TSV, then add it
to a dataset later from Reports.
Before recording starts, a dialog shows the dataset folder and the exact path
the visit will be filed under, such as
sub-01/ses-20260203/beh/sub-01_ses-20260203_task-programming_run-01_beh.tsv,
with the rule for what is added: the file and its sidecar, a row in
participants.tsv and one in that session’s scans.tsv. Recording does not
update the dataset’s combined table under
derivatives/dbs-annotator-aggregate/ entry by entry. It is refreshed once
when the recording ends: when you leave the screen, close the window, or start
or open another file. If the recording is interrupted, the table leaves out this
visit until the next merge from Reports or the next visit
recorded into the dataset refreshes it.
Whichever you choose, every entry is also written to a working copy inside the app as you record, so a crash or a closed window loses nothing. When you leave the workflow, or close the app window on Windows, macOS or Linux, the app asks whether to keep that copy or discard it. Kept, it is offered back the next time you open this workflow; discarded, it is gone and nothing is offered. It is never deleted without that answer: after a crash, or when a tablet or phone app is swiped away, the copy is always kept.
Asked when you leave the workflow or close the window.
When a copy was kept, the next time you open the workflow it is offered back:
Reopen continues the session; Discard deletes the copy. Closing the dialog any other way keeps it for next time.
Open existing TSV loads a session recorded earlier, and appends to it. The
status line under the buttons reports what was found: the row count, the next
block number, and the append number (append_id in the file: how many times
it has been reopened to add blocks), so an append is never a guess. Opening a
file also adopts the electrode models named in that file, one per side, so
the lead diagrams show the patient’s actual hardware rather than whatever the
dropdown last held. If the file names a model that is not in the catalogue, the
app says so rather than drawing the wrong lead.
A file that is not a programming session, an annotations file for instance, is refused with an explanation rather than loading as empty rows.
A session file can carry columns this app does not keep: one added by hand or by
another tool, a header name that appears twice, a value under a blank header, or a
row with more cells than the header. The app writes only the columns of the
format, so saving over such a file would delete that data. When Open existing
TSV finds any, the message names them (for example The app does not keep these
columns: lab_flag. Your file is left unchanged.) and the app never writes to
that file. You can still add blocks: every insert is saved in the app’s own
recovery copy, as for a new session that has not been exported yet, so leaving
the screen asks whether to keep that copy. A line beside Export repeats the
names of the unkept columns for as long as the file is open. Export
then saves the session without them, under a name you choose, and on the
desktop refuses the file you opened as a target, as does New. On Android
and iOS the opened file is the app’s own copy of your file, so there is nothing
to refuse. Opening another file, or starting
or reopening a session, ends this mode. Older files that spell a column the way
earlier versions did, such as block_ID or a separate date, time and
timezone, are read as before and do not count as unkept.
Step 2: Initial configuration
The state the patient arrived in, before anything is changed.
Step 2 with a baseline configuration entered: current steered across two segments of level 2 on the left lead, and the OCD clinical scale set.
Electrode model. Choose the implanted lead. The catalogue covers Medtronic, Boston Scientific, Abbott, PINS and ALEVA leads, including segmented (directional) models. The diagrams are drawn to each lead’s real contact heights and spacings, so two different leads look different. Same model on both sides is ticked by default; untick it when the two leads differ, and a second menu sets the right lead. Each side’s model is written to its own column.
Parameters, per side. Frequency, amplitude and pulse width, with quick-pick
presets. When more than one cathode is active, an amplitude split appears so you
can set the percentage per contact. Those are the E2b and E2c rows in
the screenshot above, each showing the milliamps its share works out to.
Clinical scales. The baseline assessment: disease-specific scores such as Y-BOCS or UPDRS-III. Disease presets fill the list in with a tap.
Notes. Free text. Unlike the recording step, these persist after inserting, so you can keep refining the baseline description.
Inserting records this as the baseline block (is_initial = 1).
Several cards have a gear that edits the choices they offer. These are presets: they apply to every future session and live in the app’s own preferences, never in a session file, so editing one changes nothing already recorded. The dialogs below are the ones you meet in this step.
Programs
Opened by the gear beside the Program card in steps 2 and 4.
The program labels offered by the dropdown: A, B, C
and so on, or whatever your centre uses. Stored as a preset, so the list you
build is there next session.
Parameter presets
Opened by the gear beside Parameters in steps 2 and 4.
The quick-pick chips under each stimulation field, one tab per parameter. These are shortcuts, not limits: each field accepts any value within the permitted range for that parameter, whether or not it is in this list.
A value that is not a number, or that falls outside the permitted range, is refused with the reason shown in the dialog rather than being silently dropped.
Clinical scales settings
Opened by the gear on the Clinical scales card in step 2.
Edits the disease preset buttons for the baseline assessment: the group names (OCD, MDD, PD, ET, Dystonia, TS), the scale names inside each, and each scale’s Min and Max. The range is optional and is used only to draw the longitudinal report’s clinical figure on the scale’s own axis; it is never printed. The bundled scales come with their published total ranges where a name identifies one instrument (Y-BOCS 0 to 40, MADRS 0 to 60, and so on). Names that stand for more than one, such as HAM-D, UPDRS-III or MDS-UPDRS, are left blank for you to fill in.
Selecting contacts
Tap a contact to cycle it from off, to anode, to cathode, and back to off. Tap the case to use it as the return. Segmented levels show their three segments plus a Ring strip that activates the whole level at once.
A valid configuration: two segments of level 2 as cathodes on the left, a ring contact on the right, the case as the return on both.
Invalid combinations are applied anyway and flagged as you build them, so you can work through a configuration in whatever order suits you rather than having edits rejected mid-way:
The same cathodes with no return path selected. The change is kept; the pane says why it will not stimulate.
Narrow screens
Below about 900 logical pixels (a phone, or a tablet held in portrait) the two rows stack into one column. Everything is present; there is simply more scrolling.
Step 2 in the single-column layout.
Step 3: Session scales configuration
The scale set that will be rated at every configuration.
Name the scales to be rated at every configuration, with a minimum and maximum for each. Keeping the set fixed for the whole session is what makes the ratings comparable between configurations.
Choosing a disease preset here, or having chosen one in step 2, fills the list. The gear icon edits the presets themselves, which persist between sessions. Nothing on this step is written to the file; it defines what step 4 will ask for.
Session scales settings
Opened by the gear on Session scales configuration in step 3.
The same shape as the clinical dialog, with one difference: each row here also carries a minimum and maximum. Session scales are rated on a slider at every configuration, so they need a range; clinical scales are typed once, so they do not.
Step 4: Recording
The loop, repeated once per configuration tried.
One configuration rated and ready to insert. Energy is marked not assessed.
Set the parameters and contacts as in step 2, then rate each scale. A scale that
was not assessed can be marked omitted, which writes n/a rather than a
made-up number. That is the grey bar with the crossed-out icon in the
screenshot above.
Anything observed at that configuration, a side effect included, goes in the block’s Notes, which belong to the configuration active when they were typed.
Insert to record the block. The notes clear, ready for the next one; the parameters stay, so a single amplitude change is one edit rather than a full re-entry.
Patient leaves on this, above the Insert button, marks the block being
inserted as the configuration the patient goes home with (is_final in the
file). Each export, as a report, the TSV or a BIDS dataset, asks which block
that is, defaulting to the one marked, else the last, with Not recorded as
a choice. Leaving the workflow or closing the window does not ask, so a session
that is closed without being exported records no choice. Reports call a
configuration “left on” only when it was marked this way, and “last recorded”
otherwise.
Exporting a report also asks, optionally, who recorded the session and who rated the scales. The names are printed in the report’s attestation block and saved nowhere else.
The same step in the dark theme:
Dark theme. The contact polarity colours are deliberately identical in both themes, because they are the safety-relevant part of the drawing.
Reviewing as you go
Below the entry area, everything inserted so far is shown two ways.
Four charts sharing one time axis: session scales, amplitude, pulse width and frequency.
The panels are aligned vertically, so a dip in a scale can be read against the amplitude that preceded it. They scroll horizontally, showing the most recent configurations by default, with zoom controls and drag handles to reorder the panels.
A table of every entry sits below them, grouped by block. Values that belong to the block (time, program, parameters) are printed once rather than repeated on every scale row, and a heavy rule marks each block boundary. Once scale targets are set, the rows of the best and second-best blocks are shaded in the same greens as the charts, and the shading moves as each block is inserted.
Scale targets sets what “better” means per scale (minimise, maximise, or closest to a value). Once set, the best- and second-best-scoring configurations are shaded green across all four charts. Until set, nothing is ranked; see Scale targets.
Scale targets
Opened by Scale targets in step 4, on the Reports and datasets screen, or from the report sections dialog.
Says what “better” means for each scale, which is the input the ranking needs and the one thing only you can supply. Set all: Min / Max fills the column in one tap for a set of scales that all run the same way.
Each scale gets a mode: Min, Max, Custom (closest to a stated value,
which reveals a Value field), or Ignore. Until targets are set, no
configuration is ranked anywhere: see Scale targets for the definition of
the aggregate index and What the reports do not say for its limits.
Exporting
Export offers a PDF or Word report, the raw TSV, or the whole session as a BIDS dataset. Paper size applies to both document formats and is remembered between exports.
The Export menu, with the paper-size submenu open.
Choosing a report opens the report sections dialog first. See Reports for what each section contains.
Report sections
Appears when you export a report, before the save dialog. Each report kind has its own list, and the dialog title says which one you are choosing for:
Session report sections: baseline, session scales figure, session data table, electrode configuration and programming summary.
Longitudinal report sections: clinical scales by visit, visits table, session scales by visit and block, combined session data table, electrode configuration, programming summary and source files. The session data table, the electrodes and the programming summary start unticked, because they are the sections that add pages; tick them to include them.
Each section has a one-line description of what it adds. Export is disabled while nothing is checked, because a report of a title page alone is not a document anyone wants. Scale targets… opens the dialog above without losing the selection, since the ranking those targets drive is what several of these sections show.
The session report remembers its selection for the next export. The longitudinal report opens with its defaults every time, so a section that adds pages, ticked once, is not carried into every later report.